GoLox Portable Liquid Oxygen Device GO2 finger oximeter

Specifications

Oxygen saturation display range 0% to 100%
Pulse rate display range 18 to 321 beats per minute (BPM)
Oxygen saturation declared accuracy range (Arms*) 70% to 100% SpO2 ± 2 digits
Oxygen saturation declared accuracy range (Arms*) 70% to 100% SpO2 ± 2 digits
Pulse rate declared accuracy range (Arms*) 20 to 250 BPM ± 3 digits
Low perfusion pulse rate declared accuracy range (Arms*) 40 to 240 BPM ± 3 digits
Measurement wavelengths
and output power

   Red


   Infrared



660 nanometers @ 0.8mW max average

910 nanometers @ 1.2mW max average
Temperature (operating)


Storage/transportation


Device temperature will not exceed 41°C as measured during a controlled environment test.
+41°F to +104°F
(5°C to +40°C)

-22°F to +158°F
(-30°C to +70°C)
Humidity (operating)


Storage/transportation
10% to 90% relative humidity, non-condensing

10% to 95% relative humidity, non-condensing
Operating altitude Up to 13,000 feet / 4,000 meters
Battery life (continuous)





Battery life (storage)
Approximately 2400 spot checks based on approximately 21 hours of operation using one AAA size alkaline battery, calculated at 30 seconds per use.

6 months minimum

*± 1 Arms represents approximately 68% of measurements.