GO2 finger oximeter| Oxygen saturation display range | 0% to 100% |
| Pulse rate display range | 18 to 321 beats per minute (BPM) |
| Oxygen saturation declared accuracy range (Arms*) | 70% to 100% SpO2 ± 2 digits |
| Oxygen saturation declared accuracy range (Arms*) | 70% to 100% SpO2 ± 2 digits |
| Pulse rate declared accuracy range (Arms*) | 20 to 250 BPM ± 3 digits |
| Low perfusion pulse rate declared accuracy range (Arms*) | 40 to 240 BPM ± 3 digits |
| Measurement wavelengths and output power Red Infrared |
660 nanometers @ 0.8mW max average 910 nanometers @ 1.2mW max average |
| Temperature (operating)
Storage/transportation Device temperature will not exceed 41°C as measured during a controlled environment test. |
+41°F to +104°F (5°C to +40°C) -22°F to +158°F (-30°C to +70°C) |
| Humidity (operating)
Storage/transportation |
10% to 90% relative humidity, non-condensing
10% to 95% relative humidity, non-condensing |
| Operating altitude | Up to 13,000 feet / 4,000 meters |
| Battery life (continuous)
Battery life (storage) |
Approximately 2400 spot checks based on approximately 21 hours of operation using one AAA size alkaline battery, calculated at 30 seconds per use.
6 months minimum |